“The approach of collecting sweat for CF sweat chloride tests have been established for decades. What has been a challenge has been getting enough sweat off a two-month-old, effectively, without having additional repeat visits. Building the collection path in soft microfluidics rather than rigid hardware is what let us gather more sweat, from every single participant, and still match what the lab instruments read.”
— Stephen P. Lee · Chief Technology Officer & Co-Founder, Epicore Biosystems
Study findings
The sweat chloride test has been the reference method for diagnosing cystic fibrosis since the 1950s, and the hardest patients to run it on are the ones who need it most. Infants have small bodies, delicate skin, and low sweat output. If the collection device can’t gather enough sweat, the result is invalid and the family comes back for another appointment.
This study introduces a soft, skin-interfaced sweat sticker built with microfluidic channels that draw sweat away from the skin as it’s produced, store it, and present it for colorimetric chloride analysis that can be read with a smartphone camera. The sticker conforms to the skin rather than clamping onto it, which matters for newborns.
In a comparison against the standard Macroduct sweat collection system across 51 subjects aged 2 months to 51 years, the sticker collected 26% more sweat and returned a usable sample from every participant. The standard system failed to collect enough sweat in 3 of them. Chloride values from the sticker matched measurements from clinical laboratory instrumentation, and the sticker avoided the skin indentation the standard device leaves on infants.
Scientific highlights
Collected 33% more sweat than the standard device
Across 51 subjects the sticker gathered substantially more sweat than the Macroduct sweat collection system, the current clinical standard.
A usable sample from every participant
The sticker returned enough sweat to analyze in all 51 subjects, while the standard method fell short in 3, the failure mode that sends families back for a repeat visit.
Gentler on infant skin
The soft, skin-conforming construction avoided the indentation the standard collection device presses into a newborn’s skin.
Smartphone-readable chloride results
Colorimetric analysis read by a phone camera matched chloride measurements taken on clinical laboratory instruments.
Why this research matters
A cystic fibrosis diagnosis usually starts with a positive newborn screen, and it isn’t confirmed until a quantitative sweat chloride test comes back. Every failed collection pushes that confirmation further out, and with it the start of treatment. For a family travelling to an accredited care center, a repeat visit is a real cost.
Getting enough sweat from an infant on the first attempt is the practical bottleneck in pediatric CF diagnosis. This study shows that changing the collection interface, rather than the chemistry, addresses that bottleneck directly while keeping agreement with clinical laboratory measurement.
Applications:
- Newborn screening confirmation
- Pediatric diagnostics
- Point-of-care testing
- CF care center workflow
How this connects to Epicore’s solutions
This work comes out of the long-running collaboration between Epicore Biosystems and the laboratory of John A. Rogers at Northwestern University, who co-founded Epicore. Epicore CTO Stephen P. Lee, Chief Digital Officer Alexander J. Aranyosi, and CEO Roozbeh Ghaffari are all co-authors, alongside clinicians at Lurie Children’s Hospital.
The microfluidic channel design and colorimetric readout in this sticker are the same engineering principles behind Epicore’s commercial sweat sensing platforms, which measure hydration and electrolytes for athletes and industrial workers today. This study applies them to a clinical diagnostic problem, and it sits between the foundational 2016 microfluidic sweat device and the later work on remote monitoring in adults with CF.
Connected Hydration
Real-time sweat and hydration sensing for industrial workers, with personalized intervention before heat stress sets in.
GX Sweat Patch
A wearable microfluidic patch that gives athletes sweat rate and electrolyte feedback to tailor a personalized action plan.
Discovery Patch
An FDA-registered collection system for multi-biomarker sweat analysis in clinical studies, the foundation for cortisol work.
FAQs
In the United States it starts with newborn screening. A heel-prick blood sample is tested for elevated immunoreactive trypsinogen (IRT), and depending on the state an elevated result is followed by DNA testing for CFTR mutations. A positive screen is not a diagnosis. It has to be confirmed by a sweat chloride test at a CF Foundation-accredited care center, usually with genetic testing and a clinical evaluation alongside it.
Three reasons stack up. Infants have very little skin surface to work with, their skin is delicate enough that rigid collection hardware can leave marks, and they simply produce less sweat than older children and adults. The test needs a minimum quantity to be valid, so a newborn is the most likely patient to produce an insufficient sample and need a second appointment.
It’s the collection device used in most accredited CF care centers. After pilocarpine and a weak current stimulate sweating, a rigid plastic collector held against the skin channels sweat into a spiral tube, where it’s gathered for laboratory chloride analysis. It requires at least 15 microliters of sweat to give a valid result.
The result is reported as quantity not sufficient, and the test has to be repeated. No usable answer comes back at all. In practice that means another appointment, more travel for families who don’t live near a care center, and a longer wait before treatment can start if the child does have CF.
Under Cystic Fibrosis Foundation guidance, below 30 mmol/L means CF is unlikely, 30 to 59 mmol/L is an intermediate result requiring further testing, and 60 mmol/L or above is consistent with a CF diagnosis. A clinician at a CF care center should interpret any individual result.
In this study, yes. The soft microfluidic sticker collected more sweat than the standard system and produced a usable sample from all 51 participants, where the standard method failed in 3. This was a single comparative study, though, and the standard method remains what accredited care centers use.
No. There are no needles anywhere in the procedure. A clinician applies pilocarpine to a small patch of skin and runs a weak electrical current for about 5 minutes to stimulate sweating, then collects sweat for up to 30 minutes. Babies sometimes fuss at being held still, and the area can feel warm or tingly, but the test itself isn’t painful.
About an hour for the whole appointment. Roughly 5 minutes of stimulation, up to 30 minutes of collection, and the rest in preparation and handling. The chloride measurement itself happens in a laboratory afterward, so results usually come back later rather than during the visit.
Microfluidics is the engineering of channels small enough to move tiny, precisely controlled volumes of liquid, often less than a drop. In a skin-worn device, those channels pull sweat away from the skin as it’s produced and route it to where it can be stored or measured. Working at that scale is what makes it possible to run a real chemical measurement on the small amount of sweat a person actually produces.