Skin-interfaced microfluidic biosensors for colorimetric measurements of the concentrations of ketones in sweat

Ketones, such as beta-hydroxybutyrate (BHB), are important metabolites that can be used to monitor for conditions such as diabetic ketoacidosis (DKA) and ketosis. Compared to conventional approaches that rely on samples of urine or blood evaluated using laboratory techniques, processes for monitoring of ketones in sweat using on-body sensors offer significant advantages.


Author perspective

“The interesting part of this one is how little had to change. Same passive microfluidics, same capillary valves, same camera readout as the hydration patches we manufacture. Swap the chemistry in the reservoir and you’re measuring a completely different metabolic question. That’s what a platform is supposed to mean, and it’s rare to get to demonstrate it this cleanly.”

— Alexander J. Aranyosi, PhD · Chief Digital Officer, Epicore Biosystems

Study findings

Ketones are what the body runs on when glucose is scarce. That state is deliberate on a ketogenic diet and dangerous in uncontrolled diabetes, where ketones can accumulate enough to turn the blood acidic. Either way, checking the level means a finger prick or a urine strip, and both give you one number at one moment.

This study puts the measurement on a passive skin patch. Beta-hydroxybutyrate, the most abundant ketone body, enters a microfluidic reservoir holding a paper assay. An enzyme oxidises it, which drives a coupled reaction that turns a colourless dye purple in proportion to concentration. There is no battery, no electronics and no wireless link, in deliberate contrast to the electrochemical sweat ketone systems that need all three.

The colour stabilises in about 10 minutes and reads down to 0.01 mM. Across eight subjects given ketone ester drinks, sweat ketones rose and fell with blood ketones through a dose range from 2.5 to 10 grams, with a linear correlation of R² = 0.99. Sweat concentrations run far below blood concentrations, and not by a fixed ratio: a blood level of 1.8 mM corresponded to 0.25 mM in sweat, while 3.2 mM in blood corresponded to 0.39 mM.


Scientific highlights

Sweat ketones track blood ketones

Across a 2.5 to 10 gram ketone ester dose range, sweat and blood beta-hydroxybutyrate rose together with a linear correlation of R² = 0.99.


Detects down to 0.01 mM

Sensitive enough to read the low end, with a linear response across 0 to 0.5 mM and colour that stabilises in about 10 minutes.


No electronics of any kind

A passive patch read by a camera, positioned against electrochemical sweat ketone systems that require sensors, signal electronics, wireless transmission and a battery.


Stable across real sweat conditions

Consistent readings from pH 4.4 to 8 and from 22 to 38 degrees Celsius, with negligible interference from glucose, ascorbic acid or copper.


Why this research matters

Two groups of people check their ketones regularly and for opposite reasons. People on ketogenic diets want confirmation that they’ve reached ketosis. People with diabetes need early warning of ketoacidosis, a complication that can become life-threatening. Both currently rely on episodic finger-prick or urine testing.

A patch does not solve that on its own, and this study is careful not to claim it does. What it establishes is that the signal is there and that it moves with blood, measured across a controlled dose range in healthy volunteers. Sweat ketone concentrations sit well below blood levels, roughly 0.25 mM in sweat against 1.8 mM in blood at a mid-range dose, which is why the sensitivity work matters, and why the authors describe the relationship as a suggested correlation rather than a validated substitution.



How this connects to Epicore’s solutions

This work was carried out at Northwestern University and the University of Illinois Urbana-Champaign, with Epicore’s Alexander J. Aranyosi, Donald E. Wright and Roozbeh Ghaffari among the authors, alongside Epicore co-founder John A. Rogers.

The device is architecturally the same as Epicore’s commercial patches: passive microfluidic channels driven by sweat gland pressure, capillary valves controlling flow, a colorimetric assay, and a camera for readout. Only the chemistry in the reservoir changes. That is the argument for a platform rather than a product line, and the authors make it directly, citing commercial sweat and electrolyte patches as the precedent this builds on.

Connected Hydration

Real-time sweat and hydration sensing for industrial workers, with personalized intervention before heat stress sets in.


GX Sweat Patch

A wearable microfluidic patch that gives athletes sweat rate and electrolyte feedback to tailor a personalized action plan.


Discovery Patch

An FDA-registered collection system for multi-biomarker sweat analysis in clinical studies, the foundation for cortisol work.

FAQs


Ketones are fuel molecules the liver makes from fat when glucose is in short supply, during fasting, prolonged exercise, or a very low-carbohydrate diet. The brain in particular can run on them. There are three, and beta-hydroxybutyrate is the most abundant, which is why it’s the one that gets measured.

The main ketone body in blood, usually shortened to BHB. In healthy people not fasting or dieting, blood levels sit between 0 and 0.25 mM. It rises during fasting or ketogenic eating, and rises much further in uncontrolled diabetes.

The metabolic state of running primarily on ketones rather than glucose. It happens naturally during fasting and is induced deliberately by ketogenic diets, which keep carbohydrate low enough that the body switches fuel sources. It’s a normal, regulated state, distinct from the dangerous version described below.

Three ways. A blood meter with a finger-prick strip, which is the most accurate. A urine strip, which is cheap but lags behind current blood levels. Or a breath analyser, which measures acetone rather than BHB. All three give you a reading at a single moment rather than a continuous picture.

A serious complication, most often in type 1 diabetes, where a lack of insulin drives the body to produce ketones far faster than it can use them. Enough accumulation turns the blood acidic, which is a medical emergency. Early detection matters a great deal, which is why people with diabetes are advised to check ketones when blood glucose runs high or during illness.

Yes. This study measured beta-hydroxybutyrate in sweat down to 0.01 mM using a colour-changing patch. Sweat concentrations sit well below blood concentrations, in this study around 0.25 mM in sweat when blood was at 1.8 mM, which is the main technical challenge and the reason the assay’s sensitivity is the headline result.

They move together closely in this study. Across doses from 2.5 to 10 grams of a ketone ester supplement, blood levels ranged from 0.3 to 3.2 mM and sweat from 0.096 to 0.39 mM, tracking each other with a correlation of R² = 0.99. The authors describe this as suggesting a correlation rather than establishing a substitute, and all the subjects were healthy volunteers rather than people with diabetes.

A drinkable compound that raises blood ketones within minutes without requiring fasting or a low-carbohydrate diet. It’s used in endurance sport and in research. This study used it precisely because it creates a controlled, repeatable ketone spike, which is what makes a dose-response comparison between blood and sweat possible.

Not yet, and the paper doesn’t claim it. This is early work in healthy volunteers, with no testing in people with diabetes and none at the high ketone levels that define ketoacidosis. What it establishes is that the measurement is possible and that it responds to real changes, which is the step that has to come first.